NSK Handpieces are manufactured using superior micro-precision production technology to tolerances of less than 1,000th of a millimetre & with over 100 years of experience, NSK is today regarded as the world’s largest dedicated manufacturer of quality precision high & low speed dental handpieces, micro motors & ultra-sonic scalers.

To maintain these precision instruments in peak operating condition with the efficiency & accuracy required to perform precise clinical procedures, it is recommended that all NSK products be maintained using the following Care & Maintenance Instructions.

  • NSK Care & Maintenance Instruction Sheet - Using iCare / Pana Spray Plus - PDF download link
  • NSK Care & Maintenance Instruction Sheet - Using Care3 Plus / Pana Spray Plus - PDF download link
  • NSK RECOMMENDS THE FOLLOWING CARE & MAINTENANCE PROCEDURE FOR NSK HANDPIECES

    PREPARATION

    CLEANING

    LUBRICATION

    STERILIZATION

    STORAGE

    PREPARATION
    • Contra-Angle: Remove the contra-angle from the motor and detach the bur from the chuck.
    • Air Turbine: Remove the air turbine from the coupling and detach the bur from the chuck.

    Bring instruments to a dedicated decontamination area. Remove organic contaminants with a paper wipe.

    CLEANING

    Manual: Clean the external surface of handpiece with running water (<38°C, demineralised water recommended).

    Automated NSK handpieces with this symbol are thermo-disinfector-ready and therefore may be washed and disinfected using a thermo-disinfector. When using a thermo-disinfector, be sure to follow its manufacturer’s instructions.

    LUBRICATION
    • Essential after thermo-disinfection
    • Essential prior to sterilization process

    Manual: Lubricate handpiece using NSK Pana Spray Plus / Pana Spray. To prevent mist spray from escaping into the environment, use absorbent cloth. Expel excess lubrication.

    Automated: Clean and lubricate handpieces with NSK Care3 Plus. Be sure to use the right adaptor for each handpiece.

    STERILIZATION

    Place handpieces in a sterilization pouch and seal them. EN13060 4.6.3 recommends autoclaving for 20 minutes (minimum holding time) at 121°C or 15 minutes (minimum holding time) at 132°C. NSK recommends Class B or S sterilization. (All NSK handpieces are autoclavable up to 135°C)

    STORAGE

    After the sterilization cycle, immediately remove handpieces from the autoclave. Keep them in dust-free, sterilized storage or bring them to a treatment room for use.

    Please see the video below on how to maintain your NSK products using the NSK Care3 Plus Automatic Lubrication Machine, and Pana Spray Plus Manual Lubrication Spray.